Title Extending the Functionality of Resonator Aperture Sensors for Microwave Diagnostics of Small-sized Objects
Other Titles Розширення функціональних можливостей резонаторних апертурних датчиків для НВЧ діагностики малорозмірних об'єктів
Authors Babychenko, O.Yu.
Vasiliev, Yu.S.
Galat, A.B.
Gorbenko, E.A.
Piataikina, М.I.
Shcherban, I.M.
ORCID
Keywords НВЧ діагностика
вимірювання діелектричної проникності
відкритий коаксіальний зонд
відстань між зондом та зразком
електромагнітне моделювання
microwave diagnostics
dielectric permittivity measurement
open-ended coaxial probe
tip-sample distance
electromagnetic modeling
Type Article
Date of Issue 2023
URI https://essuir.sumdu.edu.ua/handle/123456789/94062
Publisher Sumy State University
License In Copyright
Citation O.Yu. Babychenko, Yu.S. Vasiliev, A.B. Galat, et al., J. Nano- Electron. Phys. 15 No 6, 06015 (2023) DOI: https://doi.org/10.21272/jnep.15(6).06015
Abstract У роботі розглядаються основні принципи функціонування резонаторних датчиків з рухомим зондоутворюючим центральним провідником і підходи, спрямовані на оптимізацію їх вимірювальних характеристик. Показано внесок втрат, пов'язаних із виникненням хвильового поля, неузгодженістю умов поширення в обсязі резонансного вимірювального перетворювача (РВП), а також прийоми, спрямовані на виключення їх впливу. Представлені кількісні дані, що характеризують роботу сенсору, що перебудовується, в межах кожного з двох розглянутих режимів функціонування, а також переконують у необхідності оптимізації геометрії конструктиву резонаторного зонда (РЗ). Наведено порівняльний аналіз використання як РВП зондів конусного та циліндричного типів. Показано, що екранування зразків має набагато більший сенс для використання циліндричного конструктиву при діагностиці об'єктів з низьким тангенсом кута втрат. Отримана інваріантність гібридного сигналу (δQ/Q)/(δf/f) до впливу заважаючих факторів, обумовлених зміною геометрії апертурного вузла сенсора при переміщенні вістря зонда, та ступеня включення об'єкта в електромагнітне поле резонатора.
The paper discusses the basic principles of functioning of resonator sensors with a movable probeforming central conductor and approaches aimed at optimizing their measuring characteristics. The effectiveness of changing the coefficient of inclusion of an object in the electromagnetic field of the resonator is shown by introducing a mechanism into the sensor design that provides axial movement of the central conductor of the coaxial. This design allows the probing tip to be positioned both coplanar with the end aperture and displaced inside or outside the coaxial. It is shown that in order to diagnose objects with low dielectric losses, the probing tip must be pulled out of the coaxial, thereby transferring the probe to the mode of a monopole antenna. When the object has high dielectric losses, the required sensitivity of the sensor is achieved by shifting the probing tip into the inside of the coaxial, and the shortened capacitance mode is implemented. Quantitative data are presented that characterize the operation of the tunable sensor within each of the two modes of operation considered, as well as convincing of the need to optimize the geometry of the resonator probe (RP) construct. The contribution of losses associated with the appearance of a wave field, the mismatch of propagation conditions in the volume of a resonant measuring transducer (RMT), as well as techniques aimed at eliminating their influence is shown. A comparative analysis of the use of conical and cylindrical probes as RMT is presented. It is shown that the shielding of samples makes much more sense for the use of a cylindrical construct in the diagnostics of objects with a low loss tangent. It is shown that it is possible to significantly reduce the influence of interfering factors caused by a change in the geometry of the sensor aperture assembly when the probe tip is moved, by using a hybrid signal (δQ/Q)/(δf/f).
Appears in Collections: Журнал нано- та електронної фізики (Journal of nano- and electronic physics)

Views

China China
8
United States United States
15
Unknown Country Unknown Country
1

Downloads

China China
25
France France
1
United States United States
6

Files

File Size Format Downloads
Babychenko_jnep_6_2023.pdf 1,21 MB Adobe PDF 32

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.